Preventative Maintenance

  • Track and record daily N2 and He boil-off.
  • Regular lineshape tests to check shims.
  • PFG amplifier offset correction. (every 6 months)
  • Shim calibrations (monthly, or probe change)
  • PW90 checks (“regularly” to annually)
  • S/N test (every 6 months, or probe change)
  • VT calibration (as needed/six months/annual)
  • Chemical shift calibration at different temperatures (six months)
  • Sensitivity test (six months)
  • Pulse width calibration (proton, carbon, other nuclei as needed)
  • Drift test (six months)
  • Check the profile during gradient shimming. Remove probe and clean if needed.
  • Clean console air filter. (Annually or as needed according to the lab environment)
  • Check for noisy or malfunction fans and replace. (3-6 years or if they fail.
  • Test UPS battery (annual)
  • Replace UPS batteries more or less every 5 years. Some batteries can last longer.
  • Update software as needed. Discuss this with a Jeol Engineer.
  • Check lab facility air supply dryers and replace filters (3 years). Dewpoint should be at least -20C.
  •  If the instrument is showing a chronic problem, contact a service engineer to look at it.
  • Remove old data stored files from the console computer in the Remote Maintenance Tool.
Sample and testMethodDelta 6Pulse sequence
Lock: 1% CHCl3
2H lock pw906.1.2, 6.1.4, 4.12ft_2h_lock_pulse
2H homospoil cal6.1.5, 6.1.6
1H lineshape4.165.GLP Recommendft_1H_line_shape
2H homospoil selective cal.6.2
FG power adjust4.31
Water
FG power check4.3ft_pfg_z_power.jxp
0.1% EB
1H PW904.15, 4.123.GLP Recommendft_1h_90_pulse
1H sensitivity4.17, 4.131.GLP Recommendft_1H_sensitivity
1H homo SD, 1H homo Gate4.18ft_1h_sensitivity.jxp
ft_1h_homo_gate.jxp
ft_1h_homo_SD.jxp
13C ASTM no irr
13C pw904.19, 4.124.GLP Recommendft_13c_90_pulse
13C sensitivity4.20, 4.132.GLP Recommendft_13c_no_irr_sensitivity
10% EB irr
13C sensitivity4.21, 4.23ft_13c_irr_sensitivity
13C DEPT4.22ft_hp_dept_dec
Doped water 4Hz
1H homospoil selective cal.6.3? -soft pulse 2000[us]3.GLP Recommend?single_90_pulse.jxp
90 deg proton spinlock5.GLP optional1H 90 deg. Spinlock

Preventative Maintenance – Jeol