- Track and record daily N2 and He boil-off.
- Regular lineshape tests to check shims.
- PFG amplifier offset correction. (every 6 months)
- Shim calibrations (monthly, or probe change)
- PW90 checks (“regularly” to annually)
- S/N test (every 6 months, or probe change)
- VT calibration (as needed/six months/annual)
- Chemical shift calibration at different temperatures (six months)
- Sensitivity test (six months)
- Pulse width calibration (proton, carbon, other nuclei as needed)
- Drift test (six months)
- Check the profile during gradient shimming. Remove probe and clean if needed.
- Clean console air filter. (Annually or as needed according to the lab environment)
- Check for noisy or malfunction fans and replace. (3-6 years or if they fail.
- Test UPS battery (annual)
- Replace UPS batteries more or less every 5 years. Some batteries can last longer.
- Update software as needed. Discuss this with a Jeol Engineer.
- Check lab facility air supply dryers and replace filters (3 years). Dewpoint should be at least -20C.
- If the instrument is showing a chronic problem, contact a service engineer to look at it.
- Remove old data stored files from the console computer in the Remote Maintenance Tool.
Sample and test | Method | Delta 6 | Pulse sequence |
Lock: 1% CHCl3 | |||
2H lock pw90 | 6.1.2, 6.1.4, 4.12 | ft_2h_lock_pulse | |
2H homospoil cal | 6.1.5, 6.1.6 | ||
1H lineshape | 4.16 | 5.GLP Recommend | ft_1H_line_shape |
2H homospoil selective cal. | 6.2 | ||
FG power adjust | 4.31 | ||
Water | |||
FG power check | 4.3 | ft_pfg_z_power.jxp | |
0.1% EB | |||
1H PW90 | 4.15, 4.12 | 3.GLP Recommend | ft_1h_90_pulse |
1H sensitivity | 4.17, 4.13 | 1.GLP Recommend | ft_1H_sensitivity |
1H homo SD, 1H homo Gate | 4.18 | ft_1h_sensitivity.jxp ft_1h_homo_gate.jxp ft_1h_homo_SD.jxp | |
13C ASTM no irr | |||
13C pw90 | 4.19, 4.12 | 4.GLP Recommend | ft_13c_90_pulse |
13C sensitivity | 4.20, 4.13 | 2.GLP Recommend | ft_13c_no_irr_sensitivity |
10% EB irr | |||
13C sensitivity | 4.21, 4.23 | ft_13c_irr_sensitivity | |
13C DEPT | 4.22 | ft_hp_dept_dec | |
Doped water 4Hz | |||
1H homospoil selective cal. | 6.3? -soft pulse 2000[us] | 3.GLP Recommend? | single_90_pulse.jxp |
90 deg proton spinlock | 5.GLP optional1H 90 deg. Spinlock |